| Detailed Specification |
| Voltage
Range
Voltage Resolution
Voltage Regulation (Line and Load)
Voltage Accuracy
Current Sourcing
Current Limit
Current Accuracy |
0.100kV
~ 5.000kV
5V/step ± 1% + 5 V ± (1% rdg ±10V)
40mA(above 500V)
0.10 ~ 40mA, 0.02mA/step ± (1% rdg 80µ
A) |
| DC HI-POT
(GPT-715A/GPI-735A/GPI-745A) |
| Voltage Range
Voltage Resolution-Output
Voltage Regulation (Line and Load)
Voltage Accuracy
Current Sourcing
Current Limit
Current Accuracy |
0.100kV ~ 6.000kV
5V/step
1% + 5 V ± (1% rdg 10V)(above 500)
10mA(above 500V)
0.10 ~ 10mA, 0.01mA/step ± (1% rdg 50 µ
A) |
| INSULATION
RESISTANCE (GPI-725A/735A ) |
DC Voltage
Resistance Range
Resistance Accuracy
(50V/100V) Resistance Accuracy
(500V/1000V) |
0V/100V/500V/1000V
1~9900MO(500V, 1000V); 1~2000MO (50V, 100V)
1MO ~50MO :± (5% rdg ±1count)
51MO ~200MO :± (10% rdg ±1count)
201MO ~1990MO : ±(20% rdg ±1count)
1MO ~500MO :± (5% rdg ±1count)
501MO ~2000MO :± (10% rdg ±1count)
201MO ~1990MO :± (20% rdg ±1count)
2001MO ~9990MO :± (20% rdg ±1count) |
| GROUND
BOND (GPI-745A) |
Voltage Range
Current Range
Current Resolution
Current Accuracy
Ground Bond Resistance Test
Resistance Resolution
Resistance Accuracy
Test Method |
Max. 8V (DC)
3~42A
0.02A
±1% of reading ±0.2A (3A~8A); ±1%
of reading ;±50A(9A-42A)
0~620mO
0~600m (100m O)
±1% of reading ±3mO
Four terminal test mode |
| CONTINUITY
CHECK (GPT-705A/715A & GPI-725A/735A) |
Output Current Range
Compliance Voltage Range
Current Resolution |
0.1A
10%
0.1O @ 1O |
| ARC
DETECT |
| Detect Current |
40 level (1 ~ 40mA) |
| MEMORY |
| |
Groups : 10 , steps pre group : 16 |
| INTERFACE |
RS-232C
Remote I/O
Scan
GPIB
Real Panel Output |
Standard
Standard
Standard
Option
Option (GPI-745A: standard) |
| POWER
SOURCE |
| AC100V/120V/220V/230V 10%
, 50/60HZ |
| ACCESSORIES |
| User manual x 1 , GHT-105A
x 1 Power cord x 1, |
| DIMENSIONS & WEIGHT |
| 330(W) x 149(H) x mm, 446(D)
Approx. 14kg |